Basic data on CAL
NORIP Traceability
Component |
CAL Calibrator used in NORIP |
NFKK Reference Serum X |
Measurement methods |
Alanine transaminase (ALT) |
DGKC 19971 |
Transferred value from CAL in NORIP |
IFCC Reference method at 37°C |
Alkaline phosphatase (ALP) |
NORIP consensus |
NORIP, consensus |
IFCC 37°C compatible method (Roche Modular and Ortho Vitros) |
Albumin |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
RID (radial immune diffusion), Reference material: CRM 470 |
Amylase |
NORIP consensus |
NORIP, consensus |
IFCC 37°C compatible method |
Amylase, pancreatic |
NORIP consensus |
NORIP, consensus |
IFCC 37°C compatible method |
Aspartate transaminase (AST) |
NORIP consensus |
NORIP, consensus |
IFCC 37°C compatible method |
Bilirubins |
DGKC 19971 |
NORIP CAL, transferred |
Jendrassik |
Calcium |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
IDMS (isotope dilution mass spectrometry) |
Cholesterol |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
IDMS (isotope dilution mass spectrometry) |
Creatine kinase (CK) |
DGKC 19971 |
Transferred value from CAL in NORIP |
IFCC Reference method at 37°C |
Creatinine |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
IDMS (isotope dilution mass spectrometry) |
Gamma- glutamyltransferase (GT) |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
IFCC Reference method at 37°C |
Glucose |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
IDMS (isotope dilution mass spectrometry) |
HDL cholesterol |
NORIP consensus |
NORIP, consensus |
Direct methods |
Iron |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
IDMS (isotope dilution mass spectrometry) |
Lactate dehydrogenase (LD) |
NORIP consensus |
NORIP, consensus |
IFCC 37°C compatible method (Roche Modular) |
Magnesium |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
IDMS (isotope dilution mass spectrometry) |
Phosphate |
DGKC 19971 |
Transferred value from CAL in NORIP |
Ion chromatography, scintillation count |
Potassium |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
IDMS (isotope dilution mass spectrometry) |
Protein |
DGKC 19971 |
NORIP CAL, transferred |
Biuret (Doumas et. al.) |
Sodium |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Gravimetry, FAES (flame atomic emission spectrometry) |
Total iron binding capacity (TIBC)) |
NORIP consensus |
NORIP, consensus |
Transferrin methods [g/L] with IFCC calibration with factor 25.1 µmol/g to TIBC [µmol/L] |
Triglycerides |
DGKC 19971 |
Transferred value from CAL in NORIP |
GC/IDMS (gas chromatography/isotope dilution mass spectrometry) |
Urate |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
IDMS (isotope dilution mass spectrometry) |
Urea (carbamide) |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
Transferred from IMEP 17 Material 12 in Nordic Trueness Project3 |
IDMS (isotope dilution mass spectrometry) |
1. DGKC 1997
Consensus Mean and Reference Measurement values in an EQA Scheme, Uldall, A et. al., EQA News, 1998, 9 no 3
2. IMEP 17, Material 1
3. Nordic Trueness Project
NORIP data
Descriptive NORIP statistics on CAL by instrument producer (8/2-2002)
Descriptive NORIP statistics on CAL by method group (12/12-2002)
Comparison of transfer of CAL data to X in NORIP and in Nordic Trueness project (13/12-2002)